• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD MAXGUARD¿ PRESSURE RATED EXTENSION SET WITH Y-SITE(S); INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD MAXGUARD¿ PRESSURE RATED EXTENSION SET WITH Y-SITE(S); INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Catalog Number MX5301
Medical Device Problem Codes Leak/Splash (1354); Loose or Intermittent Connection (1371)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/30/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
There were multiple lot numbers reported to be involved.The information for each lot number is as follows: medical device lot #: 22099253, medical device expiration date: 13-sep-2025, and device manufacture date: 09-sep-2022.Medical device lot #: 22089181, medical device expiration date: 09-aug-2027, and device manufacture date: 08-aug-2022.Initial reporter name and address: the customer's address is unknown.New jersey, usa has been used as a default.Device evaluated by mfr: a device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the bd maxguard¿ pressure rated extension set with y-site(s) iv had a loose connection and leaked during use.This occurred once each in lots 22099253 and 22089181.The following information was provided by the initial reporter: "blue cap on the end has been changed to a clear cap and leading to connection issues, leaking of iv's, and several iv's to go bad.Was there any patient impact or patient involvement? yes, leakage around the cap and/or inability to remove cap which results in loss of iv and restick for patient.Did you experience any adverse events as a result of this reported defect? having to re insert second iv on patients".
 
Additional Manufacturer Narrative
H6: investigation summary a complaint of the new luers causing connection issues and leakage was received.A photo of the two sets was returned by the customer.Connection issues and leakage could not be confirmed from the photos received.A device history record review for model mx5301 lot number 22099253 was performed.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.A device history record review for model mx5301 lot number 22089181 was performed.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.Due to a physical sample not being received, an investigation could not be performed, and a root cause could not be determined.
 
Event or Problem Description
It was reported that the bd maxguard¿ pressure rated extension set with y-site(s) iv had a loose connection and leaked during use.This occurred once each in lots 22099253 and 22089181.The following information was provided by the initial reporter: "blue cap on the end has been changed to a clear cap and leading to connection issues, leaking of iv's, and several iv's to go bad.Was there any patient impact or patient involvement? yes, leakage around the cap and/or inability to remove cap which results in loss of iv and restick for patient.Did you experience any adverse events as a result of this reported defect? having to re insert second iv on patients".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MAXGUARD¿ PRESSURE RATED EXTENSION SET WITH Y-SITE(S)
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key16230703
Report Number9616066-2023-00049
Device Sequence Number17885625
Product Code FPA
UDI-Device Identifier10885403237553
UDI-Public10885403237553
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K051499
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue NumberMX5301
Device Lot NumberSEE H10
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/04/2023
Supplement Date Received by Manufacturer01/26/2023
Initial Report FDA Received Date01/23/2023
Supplement Report FDA Received Date02/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-