We checked the returned unit and confirmed that the ccd driver pcb fluid damage.Based on the result, we concluded that it was caused due to the fluid damage from the ccd driver pcb.In addition, we confirmed that the angle wire fluid damage, the control body fluid damage, the lg cable connector fluid damage, the control body corroded, the lg cable connector corroded, the operation channel leak, the distal body worn out, and the light guide cable bump; however, they are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|