| Model Number |
CLH-SC |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/20/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The referenced device was returned to olympus for evaluation.The evaluation confirmed the customer reported issue, even if the lamp is replaced the device does not turn on.The problem was isolated to a defective/broken temperature fuse.Additionally, the fan does not rotate.The cause of the fan malfunction is unknown; however, the investigation is still ongoing.If additional information becomes available, this report will be supplemented accordingly.In general, the end-user is required to inspect the device for defects, check the function of all devices and have alternate equipment prior to use.
|
| |
|
Event or Problem Description
|
|
Olympus (omsc) was informed via repair request that the customer halogen light source ¿does not turn on even if the lamp is replaced¿.The customer reported event was found at an unspecified event.No death or injury and no impact to patient or other has been reported to olympus.
|
| |
|
Event or Problem Description
|
|
On december 20, 2022, the customer¿s reported issue was found during inspection before use.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report was submitted to provide additional information for the event.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the definitive root cause of the thermal fuse disconnection could not be determined.Olympus will continue to monitor field performance for this device.
|
| |
|