| Catalog Number |
03P89-24 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/18/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Apoc incident #: (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
|
| |
|
Event or Problem Description
|
|
On 19-jan-2023, abbott point of care was contacted by a customer regarding i-stat pt/inr cartridges that yielded a suspected discrepant inr result of 2.9 on a patient.There was no patient information available at the time of this report.Return product is available for investigation.Method : i-stat, date: 18-jan-2023, result (units): 2.9; method : lab, date: 18-jan-2023, result (units): 0.99.Pt/inr - intended use: the i-stat pt, a prothrombin time test, is useful for monitoring patients receiving oral anticoagulation therapy such as coumadin ® or warfarin.The customer reports that the patient is not on oral anticoagulant (coumadin or warfarin) therapy.As per prothrombin time, pt/inr ifu art: 715236-00s.The customer was advised of off label use and the intended use for the i-stat pt/inr.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on limited information available that suggests the product was not performing within the variability of the assay.The investigation is underway.
|
| |
|
Event or Problem Description
|
|
Na.
|
| |
|
Additional Manufacturer Narrative
|
|
Apoc incident: # (b)(4).The investigation was completed on 29-mar-2023.A review of the device history record (dhr) confirmed the cartridge lot passed release specifications.Retained cartridge testing met the acceptance criteria in appendix 1 of q04.01.003 rev.Ak (product complaint level 2 and level 3 investigation procedure).No deficiency has been identified.
|
| |
|
Search Alerts/Recalls
|