Catalog Number 46490
Medical Device Problem Codes Fluid/Blood Leak (1250); Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/09/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is not available to be returned to the manufacturer for evaluation.The results of the investigation will be sent via a follow up medwatch.Reference: (b)(4).
 
Event or Problem Description
An end user reported an issue with a bioflo midline 5f dl-20cm max sterile barrier drape kit w/ two nitinol guidewires.It was reported that approximately 12 hours after the midline placement, the line was noted to be leaking at the catheter connection, with dual split.The line was removed and replaced bedside with a new of the same device.The second device was infused with no issues for the remainder of the patient's stay, approximately 48 hours.It was reported that there was a slight delay in the administration of a scheduled antibiotic due to the original line leaking, however this was able to be rescheduled by pharmacy once a new line was inserted and effective care was given.The patient did not experience any adverse effects or harm as a result of this incident.
 
Additional Manufacturer Narrative
As the reported device was not returned, angiodynamics is unable to perform a device evaluation.The customer's reported complaint description of midline device leaked at the junction of the extension leg tubing and the hub bifurcation cannot be confirmed since no complaint sample was returned.Without receiving product for evaluation, we are unable to definitively determine a root cause for this incident.A review of the device history records was performed for the indicated lots for any deviations related to the reported failure mode of the complaint.  the review confirms that the lots met all material, assembly and performance specifications; i.E.No ncr associated with reported failure mode.Labeling review: the directions for use (dfu), that is supplied in the reported kit contains information for the end user to reference in order to maintain the integrity of the catheter.The information that is provided includes, but is not limited to the following precautions and warnings: do not use if sterile barrier is damaged.If damage is found, call your navilyst medical representative.Inspect prior to use to verify that no damage has occurred during shipping.Do not use if package is opened or damaged.Do not fully insert catheter up to suture wing.Do not use the catheter with chemicals that are incompatible with any of its accessories, as catheter damage may occur.Acetone and polyethylene glycol-containing ointments should not be used with polyurethane catheters, as these may cause failure of the device.It is recommended that only luer lock accessories be used with the bioflo picc with endexo technology.Repeated over-tightening may reduce hub connector life.Do not use hemostats to secure or remove devices with luer lock hub connections.Catheter removal: to remove catheter, grasp catheter between suture wing and insertion site and remove slowly, in small increments, keeping catheter parallel to skin surface.Do not grasp luer lock hub to remove catheter, as catheter damage may occur.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOFLO MIDLINE
Common Device Name
MIDLINE CATHETER
Manufacturer (Section D)
ANGIODYNAMICS
10 glens falls technical park
glens falls NY 12801
Manufacturer (Section G)
ANGIODYNAMICS
10 glens falls technical park
glens falls NY 12801
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
5086587990
MDR Report Key16233059
Report Number1317056-2023-00009
Device Sequence Number15050551
Product Code FOZ
UDI-Device IdentifierH965464901
UDI-PublicH965464901
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K161866
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2024
Device Catalogue Number46490
Device Lot Number5763984
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/11/2023
Supplement Date Received by Manufacturer05/02/2023
Initial Report FDA Received Date01/24/2023
Supplement Report FDA Received Date05/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/06/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age70 YR
Patient SexMale
Patient Weight58 KG
-
-