| Model Number |
7803 |
| Medical Device Problem Codes |
Use of Device Problem (1670); Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Codes |
Fall (1848); Laceration(s) (1946); Paraplegia (2448); Respiratory Insufficiency (4462); Unintended Extubation (4564)
|
| Date of Event |
04/08/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The patient fell off the operating room table midway through the procedure causing the endotracheal tube and surgical spreaders to be ripped out of place.The doctor stated that they were tilting the bed when the patient slipped off and fell to the floor.The following morning patient was alert and oriented but paralyzed from the waist down.Unfortunately, he passed away a couple of days later due to paralysis and respiratory insufficiency.
|
| |
|
Additional Manufacturer Narrative
|
|
Patient securement to the table was insufficient for the desired degree of tilt.
|
| |
|
Event or Problem Description
|
|
Procedure was t5 to t10 laminectomy with microscopic dissection and tumor resection.Patient was prone, lower body secured to table.The table was airplaned to improve surgical access.Patient slid off the table onto the floor.
|
| |
|