• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SPINE INC THORACIC STRAIGHT PROBE; ORTHOPAEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SPINE INC THORACIC STRAIGHT PROBE; ORTHOPAEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 279702030N
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/06/2021
Type of Reportable Event Malfunction
Event or Problem Description
Voluntary report received (mw5104767) thoracic straight probe broke while in use.Both pieces retrieved.Used a broken screw removal set to remove the working end from the patient's bone.Surgery proceeded as planned.This report is for one (1) thoracic straight probe.This is report 1 of 1 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial, a follow-up will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THORACIC STRAIGHT PROBE
Common Device Name
ORTHOPAEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key16245679
Report Number1526439-2023-00134
Device Sequence Number20435226
Product Code OLO
UDI-Device Identifier10705034388328
UDI-Public(01)10705034388328
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K140927
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number279702030N
Device Catalogue Number279702030N
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/05/2023
Supplement Date Received by Manufacturer02/10/2023
Initial Report FDA Received Date01/26/2023
Supplement Report FDA Received Date02/27/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-