ZIMMER DENTAL IMPL SWISSPLUS OCTAGON 4. 1MM 8MM; DENTAL IMPLANT Back to Search Results
Model Number OPB8
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Osteolysis (2377)
Date of Event 11/25/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the implants at tooth sites #25 and #44 were removed due to bone loss around the implant.
 
Additional Manufacturer Narrative
Multiple mdr reports were filed for this event.Please see associated report: 0002023141-2023-00322.Zimmer biomet complaint number (b)(4).Weight unknown / not provided.Additional pma/510(k) number ¿k002188.A summary investigation has been completed for bone loss events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for bone loss have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of manufacturing or design defects that might lead to bone loss and escaping the available detections has been assessed and found remote and almost non-existent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).The following fields have been updated: b4: date of this report b5: describe event or problem d9: device availability g3: date received by manufacturer g6: type of report h1: type of reportable event h2: follow up type h3: device evaluated by manufacturer h6: adverse event problem h10: additional narrative dhr review was completed for the subject lot number (63721802).It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.Review completed utilizing keywords: medical: bone loss.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPL SWISSPLUS OCTAGON 4. 1MM 8MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key16246252
Report Number0002023141-2023-00324
Device Sequence Number12392870
Product Code DZE
UDI-Device Identifier00889024015821
UDI-Public(01)00889024015821(17)220831(10)63721802
Combination Product (Y/N)N
Initial Reporter CountryLE
PMA/510(K) Number
K011245
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/31/2022
Device Model NumberOPB8
Device Catalogue NumberOPB8
Device Lot Number63721802
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/06/2023
Supplement Date Received by Manufacturer04/25/2023
Initial Report FDA Received Date01/26/2023
Supplement Report FDA Received Date05/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/09/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age73 YR
Patient SexFemale
-
-