• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH 7.3MM CANNULATED LOCKING SCREW 90MM; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH 7.3MM CANNULATED LOCKING SCREW 90MM; SCREW, FIXATION, BONE Back to Search Results
Model Number 02.207.090
Medical Device Problem Codes Device Slipped (1584); Migration (4003)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 12/29/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Device report from synthes reports an event in canada as follows: it was reported that on (b)(6) 2022, a 7.3 cannulated locking screw that had been used in the second hole of the proximal end of a proximal femur lcp plate became loose and has backed out.This was noticed by the patient several days after surgery when they noticed something popping and clunking.The patient required hardware removal or revision surgery.No further information is available.This report involves one 7.3mm cannulated locking screw 90mm.This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: additional device product codes: hrs, hty, jdw.Complainant part is not expected to be returned for manufacturer review/investigation.Initial reporter is a j&j employee.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h6: a manufacturing record evaluation cannot be performed since the lot number is unknown.The photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the device from the received image(s).Visual analysis of the photo revealed that the 7.3mm cannulated locking screw 90mm had moved from its original positioning.X-ray image provided shows a plate and screw construct at the left proximal femur.Migration of the second most proximal screw can be confirmed since it almost protrudes from the skin and the head is no longer seated at the insertion hole of the plate, the screw has backed out more than halfway through.Functionality of the device cannot be assessed through photo investigation, therefore this condition cannot be confirmed.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for the 7.3mm cannulated locking screw 90mm.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
7.3MM CANNULATED LOCKING SCREW 90MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key16247764
Report Number8030965-2023-01003
Device Sequence Number17780231
Product Code HWC
UDI-Device Identifier10886982051875
UDI-Public10886982051875
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K000066
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number02.207.090
Device Catalogue Number02.207.090
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/08/2023
Supplement Date Received by Manufacturer02/01/2023
Initial Report FDA Received Date01/26/2023
Supplement Report FDA Received Date02/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK - PLATES: LCP PROXIMAL FEMUR
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-