| Model Number |
OPB10 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Pain (1994); Osteolysis (2377)
|
| Date of Event |
11/29/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the implant at tooth 35 (fdi) was removed due to bone loss around the implant.Pain was reported as a result of the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimvie complaint: (b)(4).Patient weight is not provided / unknown.Initial reporter¿s title is not provided / unknown.Additional 510(k) number is k002188.A summary investigation has been completed for bone loss events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for bone loss have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of manufacturing or design defects that might lead to bone loss and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint number (b)(4).The following fields have been updated: b4: date of this report.B5: describe event or problem.D4: additional device information.D9: device availability.G3: date received by manufacturer.G6: type of report.H1: type of reportable event.H2: follow up type.H3: device evaluated by manufacturer.H4: device manufacturer date.H6: adverse event problem.H10: additional narrative.Dhr review was completed for the subject lot number (63796497).It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Complaint history review was performed for the reported lot number (63796497) for similar events and one (1) other complaint was identified.Investigation has yet to be completed and results are unavailable at this time.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|
Search Alerts/Recalls
|