| Model Number |
E2515H |
| Medical Device Problem Codes |
Arcing (2583); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Burn(s) (1757)
|
| Date of Event |
01/06/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during tracheostomy, while making an incision to perform the procedure, the sparks flew from the pencil and made a small fire with the surrounding fabric that caused the burn around patient's neck.There was a temporary injury which was a micro burns on skin around the patient's neck with the size of 1 centimeter.Photos were submitted and the tip of the pencil had aburn mark.There was no need for medical or surgical intervention as well as extended hospitalization.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5, e1 (first and last name, phone number), e3, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.Visual inspection noted the device electrode appears to have eschar on the tip.It was reported that the use of the device resulted in a burn and the device discharged a visible spark or arc.There was an operating room fire without injury.The reported issues could not be confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during tracheostomy, while making an incision to perform the procedure, the sparks flew from the pencil and made a small fire with the surrounding fabric that caused the burn around patient's neck.The small fire was extinguished by covering with another cloth.There was a temporary injury which was a microburns on skin around the patient's neck with the size of 1 centimeter.The procedure was completed using a non-medtronic product.Photos were submitted and the tip of the pencil had a burn mark.There was no need for medical or surgical intervention as well as extended hospitalization.
|
| |
|