MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERPICC PROVENA SOLO 5F DL FT MAX BARRIER PLUS KIT; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERPICC PROVENA SOLO 5F DL FT MAX BARRIER PLUS KIT; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Medical Device Problem Code Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/06/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The manufacturer has received the sample and is pending evaluation.Results are expected soon.A lot history review (lhr) of regs0826 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported by the director of cath lab that a long coiled hair was found within a sterilized picc kit.It was stated the kit was not used on a patient.No other information was provided.
 
Event or Problem Description
It was reported by the director of cath lab that a long coiled hair was found within a sterilized picc kit.It was stated the kit was not used on a patient.No other information was provided.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of a fiber in the kit was inconclusive due to the sample condition.Two photographs of a previously opened 5fr d/l powerpicc solo ft catheter kit were returned for evaluation.The kit components appeared unused as most components were within their respective protective packaging sheaths and no residual material from use was noted on the components.A long dark curled fiber was noted in the procedural tray, near the chloraprep and guidewire.The fiber appeared consistent with a hair or hair-like material.As the kit had been opened, it could not be independently confirmed if the fiber had been packaged in the kit or if it was introduced into the tray after opening the tray.A review of the manufacture quality and inspection records for the specific lot involved found no potentially related issues encountered during the manufacture and assembly of that product lot.Manufacturing controls are in place to mitigate the occurrence of this type of failure.The report of a foreign material within the packaging has been documented and included in complaint trending.This type of complaint will continue to be monitored as part of ongoing efforts to identify potential manufacturing related issues.H3 other text: evaluation findings are in section h11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC PROVENA SOLO 5F DL FT MAX BARRIER PLUS KIT
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
kayla olsen
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key16250119
Report Number3006260740-2023-00139
Device Sequence Number6406320
Product Code LJS
UDI-Device Identifier00801741155215
UDI-Public(01)00801741155215
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K072230
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2023
Device Model NumberN/A
Device Catalogue NumberS1295108FD5
Device Lot NumberREGS0826
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/19/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/06/2023
Supplement Date Received by Manufacturer05/12/2023
Initial Report FDA Received Date01/26/2023
Supplement Report FDA Received Date05/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-