MAUDE Adverse Event Report: ZIMMER DENTAL IMP TM 4.1MM MTX FULL, 11.5MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER DENTAL IMP TM 4.1MM MTX FULL, 11.5MM; DENTAL IMPLANT Back to Search Results
Model Number TMT4B11
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/30/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimmer biomet complaint number: (b)(4).Age and date of birth unknown / not provided.Weight unknown / not provided.Pma/510(k) numbers: k113753, k112160.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the implant on tooth site #19 became broken in half and was removed.
 
Additional Manufacturer Narrative
This report is being submitted to report additional information.The product was returned, and the reported event was confirmed following visual evaluation.Based on the evaluation, device malfunction has occurred.However, there is no existing nonconformance / capa / hhe/d / ie / product hold against the reported device that did or could cause or contribute to the reported event.Monthly post market trending review identified no adverse trends or actionable trends for the reported event or device.Zimvie quality management system (qms) has controls in place to ensure the distribution of conforming products.Therefore, based on the available information, the product was within specifications and conforming when it left zimvie.Dhr review for the lot had revealed no deviations nor non-conformances which could have caused or contributed to the reported events.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.A definitive root cause could not be determined.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimvie will continue to monitor for trends.
 
Event or Problem Description
No additional or corrected information to report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMP TM 4.1MM MTX FULL, 11.5MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key16250730
Report Number0002023141-2023-00343
Device Sequence Number6605817
Product Code DZE
UDI-Device Identifier00889024018945
UDI-Public(01)00889024018945(17)240824(10)1231594
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K132258
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/24/2024
Device Model NumberTMT4B11
Device Catalogue NumberTMT4B11
Device Lot Number1231594
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/03/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/03/2023
Supplement Date Received by Manufacturer04/24/2023
Initial Report FDA Received Date01/26/2023
Supplement Report FDA Received Date05/15/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/26/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-