| Model Number |
328466 |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/25/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medical device expiration date: unknown.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
|
| |
|
Event or Problem Description
|
|
It was reported that the bd insulin syringe with the bd ultra-fine¿ needle pierced through the shield after the consumer recapped it.This occurred with 2 needles.The following information was provided by the initial reporter: "consumer reported gest diabetic reshielded syringe after injection and stuck her finger when needle went thru needle shield.2 times, 2 different syringes from this box.Advised caller to not reshield the syringes".
|
| |
|
Event or Problem Description
|
|
It was reported that the bd insulin syringe with the bd ultra-fine¿ needle pierced through the shield after the consumer recapped it.This occurred with 2 needles.The following information was provided by the initial reporter: "consumer reported gest diabetic reshielded syringe after injection and stuck her finger when needle went thru needle shield.2 times, 2 different syringes from this box.Advised caller to not reshield the syringes".
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary.No samples including photos were returned therefore the complaint could not be confirmed and the root cause is undetermined.Due to the batch being unknown, no dhr review can be completed.
|
| |
|
Additional Manufacturer Narrative
|
|
The following fields were updated due to additional information: d10: device available for eval yes, d10: returned to manufacturer on: h6: investigation summary customer returned 23 used syringes 0.5ml 30ga with no polybag.It was reported by the consumer that needle went through shield and cause needle stick twice.All the syringes were visually inspected and was verified that only one of the returned syringes has the needle sticking out through the shield.No issues were found on other 22 syringes.Needle stick could have happened because the cannula was through the shield.Due to the batch being unknown, no dhr review can be completed.Based on the samples received, embecta was able to confirm the customer¿s indicated failure.No root cause can be determined at this time.
|
| |
|
Event or Problem Description
|
|
It was reported that the bd insulin syringe with the bd ultra-fine¿ needle pierced through the shield after the consumer recapped it.This occurred with 2 needles.The following information was provided by the initial reporter: "consumer reported gest diabetic reshielded syringe after injection and stuck her finger when needle went thru needle shield.2 times, 2 different syringes from this box.Advised caller to not reshield the syringes".
|
| |
|