|
No complaint was received with the return of the device.Failure event observed during analysis.Final analysis found premature battery depletion.No sources of high current were noted.The cause of the premature battery depletion was consistent with lithium (li) cluster formation.From these analyses, in the absence of high current draw, it is probable that the premature battery depletion was caused by a lithium cluster induced short circuit.Li clusters are a known depletion mechanism for these advisory products that has been investigated and associated with a field action in october 2016.
|