| Model Number |
MI1000 MED-EL CONCERT |
| Medical Device Problem Codes |
Impedance Problem (2950); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The user's hearing performance with the device has deteriorated after a head trauma occurred.Re-implantation surgery is scheduled in (b)(6) 2023.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance with the device has deteriorated after a head trauma which occurred in (b)(6) 2020.The user_s performance with the device was good before the event occurred.Re-implantation has been done on (b)(6)2023.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusions: damage to the active electrode which is consistent with minute device mobility was determined to have led to device failure over time due to fatigue breakages.Also, an impact to the head might have weakened the fixation of the active electrode and could also be a factor for electrode mobility.The investigation results appear to match the problems mentioned in the recipient report.This is a final report.
|
| |
|