• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONVATEC DOMINICAN REPUBLIC INC ESTEEM+; POUCH, COLOSTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONVATEC DOMINICAN REPUBLIC INC ESTEEM+; POUCH, COLOSTOMY Back to Search Results
Model Number 416740
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
The end user's caregiver reported that she inspected two market units with same lot and all twenty pouches in the boxes appeared with the starter hole off centered.She did not allow the end user to wear any of these pouches.The photographs were received from the complainant depicting the issue.
 
Additional Manufacturer Narrative
Device 7 of 20.Based on the available information, this event is deemed to be a reportable malfunction.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.(b)(4).
 
Additional Manufacturer Narrative
Additional information - this emdr is being submitted to include the below: h6: investigation results under imdrf cause investigation code, imdrf investigation findings, imdrf cause conclusions h10: investigation summary photograph, video and/or physical sample evaluation: there are photographs associated with this case and in these, the reported defect can be seen.No unused return sample was expected.Batch record revision results: lot 1k00313j was manufactured on 10/6/2021, in convex 1 pc line, with a total of (b)(4) mkus.The compliance engineer performed a batch record review on 02/oct/2023, to verify if all the applicable procedures were followed and no issues were found; all the components for assembly were correct per bill of materials (bom) and all the tooling information documented was also correct, system application product (sap) material id 1703550 and manufacturing order (b)(4).The production process, in-process control, testing results and packaging of products was run according to the process instruction (pi) and recorded in br31-146 ver.7.0.Review of the batch record showed that all relevant tests required during the manufacturing process and final product release had been fulfilled and met the requirements.No discrepancy related to this issue were found within the documentation.Historical complaints review: on 02/oct/2023, compliance engineer ran a query in database in order to verify the complaints reported for the 1k00313j lot for the malfunction code ¿skin barrier starter hole is defective (e.G misalignment or off center), leakage may occur (pre-cut only)¿ and as result, no additional type 2 complaints were identified during this search as per work instruction.Historical nonconformance review: on 02/oct/2023, compliance engineer ran a query in database looking for any in process nonconformance / corrective action/preventive action (cap)a (s) associated to the malfunction code ¿skin barrier starter hole is defective (e.G misalignment or off center), leakage may occur (pre-cut only)¿ for the lot number 1k00313j and as result, no nonconformance / capa (s) for this malfunction code were generated during the manufacturing process of the referenced lot.Current quality controls: based on the process instruction, the following tests are performed in the manufacturing lines, in order to identify this failure mode in our manufacturing process: pd31-146 ¿dimensional wafer ctq¿: ¿ frequency: 16 pouches per turn (2 pouches per hour) ¿ sample quantity: 16 pouches per turn (first piece at start) ¿ acceptance criteria: accept = 0 | reject = 1 defect rate analysis: there have only been 20 defective part confirmed to date from a lot size of 8480 products.This represents a defect rate of only 0.24%, which is well within an appropriate acceptable quality level (aql) for this defect which should be 2.5% based on our pd31-146 ver.2.0 ¿convex 1 pc sampling plan¿.In addition, all of the in-process testing on this lot did not find a single defective pouch, which confirms that the lot is unlikely to breach an aql of 2.5.This issue certainly appears to be an isolated incident, but more importantly to date, it is well within our accepted aql level for this type of failure mode or defect.Conclusions: the review of the batch record for lot 1k00313j showed that all relevant tests required during the manufacturing process and final product release had been fulfilled and met the requirements, all applicable manufacturing and quality processes were followed, and no discrepancies or deviations were recorded.No changes to the end-to-end manufacturing process or components used during assembly of the batch were made.No nonconformity has been registered during the manufacturing process of the affected lot for the malfunction code ¿skin barrier starter hole is defective (e.G misalignment or off center), leakage may occur (pre-cut only)¿.No additional complaints were reported for lot affected related to the malfunction code ¿skin barrier starter hole is defective (e.G misalignment or off center), leakage may occur (pre-cut only)¿.Based on this, no negative trend was identified.Based on preliminary investigation results, there is no objective evidence that other products from this lot are impacted, and the issue appears to be isolated.This issue will be monitored through the post market product monitoring review process.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number reporting site: 1049092 manufacturing site: 9618003.
 
Event or Problem Description
To date no additional patient or event details have been received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ESTEEM+
Common Device Name
POUCH, COLOSTOMY
Manufacturer (Section D)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal 91000
DR  91000
Manufacturer (Section G)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal 91000
DR   91000
Manufacturer Contact
pamela meadows
7815 national service road
suite 600
greensboro, NC 
3365424679
MDR Report Key16252168
Report Number9618003-2023-00191
Device Sequence Number5682499
Product Code EZQ
UDI-Device Identifier00768455188499
UDI-Public00768455188499
Combination Product (Y/N)N
Initial Reporter StateIN
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date10/05/2026
Device Model Number416740
Device Lot Number1K00313J
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/04/2023
Supplement Date Received by Manufacturer10/04/2023
Initial Report FDA Received Date01/27/2023
Supplement Report FDA Received Date10/26/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/06/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-