| Model Number |
1219-32-148 |
| Medical Device Problem Codes |
Device Dislodged or Dislocated (2923); Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Pain (1994); Insufficient Information (4580)
|
| Date of Event |
01/10/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Surgeon sent an email directly to complaint handling unit: he reported about an marathon inaly damage.Revision of inlay component on (b)(6) 2023.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : the device associated with this report was returned to depuy synthes for evaluation.Visual examination of the device found the interlock feature fractured.Deformation was also found on the edge of the liner, it is evident that it was loaded by the femoral head and patient body weight.Additionally, machining lines from the manufacturing process were yet visible within the liner which would not be as obviously present had the femoral head been more centrally loaded.Based on the observations, it is reasonable to conclude that a disassociation event occurred between the liner and the cup.However, no signs of wear or deterioration were found.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : 1) quantity manufactured: (b)(4).2) date of manufacture: 2020/08/13 3) any anomalies or deviations identified in dhr: there were no non-conformances associated with this lot.4) expiry date: 2025/07/31 5) ifu reference: ifu-0902-00-701.Device history review : a manufacturing record evaluation was performed for the finished device [121932148 / j8850z] number, and no non-conformances / manufacturing irregularities were identified during manufacturing.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint#: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: h6 (clinical code).
|
| |
|
Event or Problem Description
|
|
Additional information received indicated that there was no surgical delay reported.The patient noticed "cracking" at the area of the right hip since about 4 months.The pain extends to the spine.No known trauma.After implantation of a right hip.
|
| |
|
Search Alerts/Recalls
|