Model Number 1219-32-148
Medical Device Problem Codes Device Dislodged or Dislocated (2923); Naturally Worn (2988)
Health Effect - Clinical Codes Pain (1994); Insufficient Information (4580)
Date of Event 01/10/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Surgeon sent an email directly to complaint handling unit: he reported about an marathon inaly damage.Revision of inlay component on (b)(6) 2023.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : the device associated with this report was returned to depuy synthes for evaluation.Visual examination of the device found the interlock feature fractured.Deformation was also found on the edge of the liner, it is evident that it was loaded by the femoral head and patient body weight.Additionally, machining lines from the manufacturing process were yet visible within the liner which would not be as obviously present had the femoral head been more centrally loaded.Based on the observations, it is reasonable to conclude that a disassociation event occurred between the liner and the cup.However, no signs of wear or deterioration were found.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : 1) quantity manufactured: (b)(4).2) date of manufacture: 2020/08/13 3) any anomalies or deviations identified in dhr: there were no non-conformances associated with this lot.4) expiry date: 2025/07/31 5) ifu reference: ifu-0902-00-701.Device history review : a manufacturing record evaluation was performed for the finished device [121932148 / j8850z] number, and no non-conformances / manufacturing irregularities were identified during manufacturing.
 
Additional Manufacturer Narrative
Product complaint#: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: h6 (clinical code).
 
Event or Problem Description
Additional information received indicated that there was no surgical delay reported.The patient noticed "cracking" at the area of the right hip since about 4 months.The pain extends to the spine.No known trauma.After implantation of a right hip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN MAR +4 10D 32IDX48OD
Common Device Name
PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key16252686
Report Number1818910-2023-02176
Device Sequence Number5400064
Product Code LPH
UDI-Device Identifier10603295014034
UDI-Public10603295014034
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K033273
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1219-32-148
Device Catalogue Number121932148
Device Lot NumberJ8850Z
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/16/2023
Supplement Date Received by Manufacturer02/20/2023
03/30/2023
Initial Report FDA Received Date01/27/2023
Supplement Report FDA Received Date02/21/2023
03/31/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/13/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
APEX HOLE ELIM POSITIVE STOP.; CORAIL2 STD SIZE 10.; HIP BALL BIOLOX 32 +1 12/14.; PINN 100 W/GRIPTION 48MM.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age73 YR
Patient SexFemale
-
-