| Model Number |
PREVUE C (+18.00 D) |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
12/21/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This initial emdr is being submitted to fda as a reportable event that occurred in the usa.Haptic damage is indicated as a potential malfunction related to the iol, as covered under the warnings section of the product's instructions for use (ifu).Rmanufacturer's codes for: type of investigation, findings, and conclusion are pending device return and completion of product investigation.Once the product investigation is completed, a follow-up report will be submitted to fda which will include the manufacturer's codes for type of investigation, findings, and conclusion.
|
| |
|
Event or Problem Description
|
|
Damaged haptic after implantation lens haptic broke after implantatio; back haptic tore off on insertion remained in shooter.They explanted the lens with the incision enlarged and no sutures required.Patient impact: device explantation patient current prognosis: outcome is good.
|
| |
|
Event or Problem Description
|
|
Damaged haptic after implantation.Lens haptic broke after implantation; back haptic tore off on insertion remained in shooter.They explanted the lens with the incision enlarged and no sutures required.Patient impact: device explantation patient current prognosis: outcome is good.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up #1 emdr is being submitted to fda for a reportable event that occurred in the usa.Note: delayed emdr submission was due to fda esg webtrader account system software issues, per fda esg help desk ticket # (b)(4) (and subsequent ticket # (b)(4)).Hoya due diligence is documented under internal (b)(4).The report includes corrected information and additional information not available/included in the initial report.The product was returned to the manufacturer.The investigation was conducted, with the methods and results as noted below.Appearance check result was consistent to reported information.No abnormalities were found in production and inspection records of the product.(serial no.: (b)(6); model: prevue c).We also confirmed there were not any abnormalities on the loop pull strength test record of the material lot.(soth-g115-03).The dye test result showed the injector tip was properly coated.Proper coating allows the lens to advance.We could release a re-installed iol from the returned injector without any problems.The exact root cause of the event was not determined.However, based on available information, we believe this event was not caused by our product quality.A review of the most recent complaint trending data indicates that no significant trends have been identified at this time and no capa is required as part of the product evaluation.
|
| |
|