• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US TRI-LOCK BPS SZ 4 STD OFFSET; TRILOCK HIP STEM : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US TRI-LOCK BPS SZ 4 STD OFFSET; TRILOCK HIP STEM : HIP FEMORAL STEM Back to Search Results
Model Number 1012-04-040
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Hypersensitivity/Allergic reaction (1907); Pain (1994); Pocket Erosion (2013); Metal Related Pathology (4530)
Date of Event 01/17/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Patient presented in surgeon's clinic with a painful right hip.Upon examination and studies surgeon elected to revise the hip since it contained an asr construct.Upon more lab work it was determined the patient had elevated metal ion levels which surgeon attributed to the metal-on-metal index construct.A mri study also showed a pseudo tumor index surgery was performed on (b)(6) 2009 at (b)(6) hospital in (b)(6) by another surgeon.Surgeon performed a right hip revision surgery on tuesday, (b)(6) 2023 at (b)(6) hospital.Upon exposure to the joint space surgeon used a punch to disengage the femoral head construct.Both the head and neck spacer came disengaged together as a construct from the trunion.Surgeon then used a competitor¿s cup out system to break the boney in growth of the asr acetabular shell which he successfully did.Surgeon then elected to use a competitor¿s acetabular shell and acetabular liner to implants.Surgeon determined the trilock bps sz 4 std stem was well fixed and elected to keep it implanted.Surgeon then trialed the construct with depuy femoral head trials until he found the construct he liked.No index components were deemed loose.No instrumentation broke during the procedure.There were no boney fractures or soft tissue tears.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: a manufacturing record evaluation (mre), was not possible because the required lot code was not valid.
 
Additional Manufacturer Narrative
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Medical records received.After review of the medical records the patient was revised to address metallosis, pseudotumor and persistent pain.Operative note reported there was trunnionosis that was debrided in the trunnion however in good shape.There was a large pseudotumor, substantial evidence of metallosis inside the femoral head this was debrided.There was soft tissue erosion but abductors were intact.Doi: (b)(6) 2009; dor: (b)(6) 2023; right hip.
 
Additional Manufacturer Narrative
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Medical records received and stated that the operative noted that there was considerable soft tissue erosion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRI-LOCK BPS SZ 4 STD OFFSET
Common Device Name
TRILOCK HIP STEM : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key16253601
Report Number1818910-2023-02227
Device Sequence Number5472992
Product Code LPH
UDI-Device Identifier10603295000754
UDI-Public10603295000754
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P070026
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 01/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1012-04-040
Device Catalogue Number101204040
Device Lot NumberC3KAR1000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/17/2023
Supplement Date Received by Manufacturer02/23/2023
12/26/2023
12/27/2023
Initial Report FDA Received Date01/27/2023
Supplement Report FDA Received Date02/24/2023
12/26/2023
12/29/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ADAPTER SLEEVES 12/14 +8; ASR ACETABULAR CUPS 58; ASR UNI FEMORAL IMPL SIZE 51; TRI-LOCK BPS SZ 4 STD OFFSET
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
Patient SexMale
-
-