| Model Number |
5820 |
| Medical Device Problem Codes |
Premature Discharge of Battery (1057); Failure to Interrogate (1332); Pacemaker Found in Back-Up Mode (1440); No Pacing (3268)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
During an in clinic follow up, it was noted the device was found in back up mode and unable to be interrogated.Technical support was contacted and suspected the battery was drained.A magnet was attempted and there was no response from the device.The device was explanted and replaced to resolve the event.The patient was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Implant date is estimated to have occurred in 2015.Further information was requested but not received.
|
| |
|