• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVANOS MEDICAL INC. MICROCUFF PEDIATRIC ENDOTRACHEAL TUBE; VAP MICROCUFF ENDOTRACHEAL TUBES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. MICROCUFF PEDIATRIC ENDOTRACHEAL TUBE; VAP MICROCUFF ENDOTRACHEAL TUBES Back to Search Results
Model Number 35111
Medical Device Problem Code Separation Problem (4043)
Health Effect - Clinical Codes Hypoventilation (1916); Hypoxia (1918)
Date of Event 01/04/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The actual complaint product was not returned for evaluation.A review of the device history record is in-progress.All information reasonably known as of 25 jan 2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the reported event but is relaying the information provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.Device not returned.
 
Event or Problem Description
It was reported that the top of the endotracheal tube came away from the adapter.While disconnecting to initiate high-frequency oscillatory ventilation (hfov) from conventional, the tip disconnected from the patient resulting in the patient not being ventilated, losing oxygenation, and dropped oxygen saturation.No medical interventions reported.Per additional information received on 17 jan 2023, the patient¿s oxygen saturation decreased to 25%.Fi02 was increased.The device was in use for approximately 48 hours.The patient has recovered from the incident.
 
Additional Manufacturer Narrative
The device history record for lot cm2154001 was reviewed, and the product was produced according to product specifications.It is not possible to determine the root cause of the reported issue without a sample, as functional evaluation could not be performed to confirm or duplicate the reported incident.All information reasonably known as of 09 mar 2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the reported event but is relaying the information provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Additional Manufacturer Narrative
] additional information was requested.All information reasonably known as of 10 apr 2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the reported event but is relaying the information provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICROCUFF PEDIATRIC ENDOTRACHEAL TUBE
Common Device Name
VAP MICROCUFF ENDOTRACHEAL TUBES
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
COMMAND MEDICAL NICARAGUA SA
km 12.5 carretera norte
parque industrial las mercedes, edificio 16
managua managua, ni
Manufacturer Contact
lisa clark
5405 windward parkway
alpharetta, GA 30004
4704485444
MDR Report Key16257413
Report Number3011270181-2023-00006
Device Sequence Number5400000
Product Code BTR
UDI-Device Identifier00609038351117
UDI-Public00609038351117
Combination Product (Y/N)N
Initial Reporter CountryEI
PMA/510(K) Number
K080821
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number35111
Device Catalogue NumberN/A
Device Lot NumberCM2154001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/06/2023
Supplement Date Received by Manufacturer03/06/2023
03/06/2023
Initial Report FDA Received Date01/27/2023
Supplement Report FDA Received Date03/13/2023
04/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/03/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
Patient Weight3 KG
-
-