| Model Number |
75100478 |
| Medical Device Problem Codes |
Fracture (1260); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Bone Fracture(s) (1870); Failure of Implant (1924)
|
| Date of Event |
03/19/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that, after a left surgery performed on (b)(6) 2013 to address osteoarthritis symptoms, the patient sustained peri-prosthetic fracture of the stem that made necessary a revision surgery on (b)(6) 2019.During this procedure, the femoral stem, modular head were exchanged with competitor devices.This information was provided by the national joint registry of the united kingdom, as part of a retrospective data collection of patients who underwent a primary thr surgery with a hip prosthesis construct that included a polarstem stem with an r3 acetabular cup and that required a revision surgery due to specific reasons.As such, no further information will be available.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).This complaint was opened by smith+nephew to document a patient complication identified through a review of the national joint registry uk from united kingdom that includes reference to the use of a smith+nephew product.The reported issue(s) relate to known inherent procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to confirm a relationship between the reported event and the device or identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: health effect and medical device problem codes updated.
|
| |
|
Search Alerts/Recalls
|