| Model Number |
84591 |
| Medical Device Problem Codes |
Break (1069); Defective Device (2588); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/30/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the catheter was fractured.A direxion hi-flo fathom-16 system was selected for use.During the procedure, advancement of coils was difficult due to microcatheter stretching towards microcatheter tip.It was noted that the catheter fractured during attempted advancement and withdrawal, within the base catheter, with just mild to moderate force.The procedure was completed with a different device.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: the device was returned for evaluation.The device shaft was microscopically analyzed for any damage.The guidewire was stuck in the device.The device showed a fractured shaft located 40.2cm from the hub.The device was not completely separated, the inner liner was still intact.The inner liner was stretched and constricting the guidewire from movement.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.No other issues were identified during the product analysis.
|
| |
|
Event or Problem Description
|
|
It was reported that the catheter was fractured.A direxion hi-flo fathom-16 system was selected for use.During the procedure, advancement of coils was difficult due to microcatheter stretching towards microcatheter tip.It was noted that the catheter fractured during attempted advancement and withdrawal within base catheter with just mild to moderate force.The procedure was completed with a different device.No patient complications were reported.
|
| |
|
Search Alerts/Recalls
|