• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDINAL HEALTH DOBBHOFF 10FR 43IN W STYLET EN; TUBES, GASTROINTESTINAL (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH DOBBHOFF 10FR 43IN W STYLET EN; TUBES, GASTROINTESTINAL (AND ACCESSORIES) Back to Search Results
Model Number 8884711006E
Medical Device Problem Codes Difficult to Remove (1528); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/19/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event or Problem Description
The customer reported that when the nurse unscrewed the feeding tube from the catheter, the connector tip came off of the feeding tube.The nurse had to use force to unscrew the feeding tube which was stuck in the connector.Per additional information received, the nurse put the device back in and the tube was still permeable and functional.There was no impact to the patient.
 
Additional Manufacturer Narrative
Section h6.Medical device problem code for ¿difficult to remove¿ was selected in error.
 
Additional Manufacturer Narrative
The device history record was reviewed and indicated that the product was released accomplishing all quality standards.The device was manufactured on august 19, 2020.The device was not received for evaluation, but five photos were provided, and the reported condition has been confirmed.A corrective and preventative action has been initiated to address the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DOBBHOFF 10FR 43IN W STYLET EN
Common Device Name
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX   22500
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key16259156
Report Number9612030-2023-03538
Device Sequence Number17338125
Product Code KNT
UDI-Device Identifier10884521582903
UDI-Public10884521582903
Combination Product (Y/N)N
Initial Reporter CountryCA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8884711006E
Device Catalogue Number8884711006E
Device Lot Number2022314964
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/12/2023
Supplement Date Received by Manufacturer01/12/2023
01/12/2023
Initial Report FDA Received Date01/27/2023
Supplement Report FDA Received Date02/16/2023
03/02/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/19/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-