| Model Number |
4408000000 |
| Medical Device Problem Code |
Material Disintegration (1177)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/31/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Reported events: 1 event was reported for this quarter.Product return status: 1 device investigation type has not yet been determined.Additional information: 1 device was not labeled for single-use.1 device was not reprocessed or reused.
|
| |
|
Event or Problem Description
|
|
This report summarizes 1 malfunction event in which the device was shedding metal debris.- 1 event had no patient involvement; no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Supplemental rationale 1 previously reported event is included in this follow-up record.Product return status 1 device was received.
|
| |
|
Event or Problem Description
|
|
This report summarizes 1 malfunction event in which the device was shedding metal debris.1 event had no patient involvement; no patient impact.
|
| |
|
Search Alerts/Recalls
|