| Model Number |
4.75MM BC KNOTLESS SWIVELOCK |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Device Embedded In Tissue or Plaque (3165); Insufficient Information (4580)
|
| Date of Event |
01/09/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
On 1/10/2023, it was reported by a sales representative via email that an ar-2324kbcc biocomposite knotless swivelock anchor came out without the eyelet after implantation.An ar-2324bcc biocomposite swivelock failed during insertion and the eyelet was left in the socket.An ar-3632sp fibertak sp inserter bent during insertion.This was discovered during an arthroscopic rotator cuff repair procedure on (b)(6) 2023.
|
| |
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The most likely cause for the reported failure is use error due to excessive force on the device.
|
| |
|