Model Number Ø9X 20MM BC IF SCRW, VENTED
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/09/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
It was reported that during an anterior cruciate ligament surgery during inserting the screw the distal tip broke off and prevented the implantation.This incident occurred with a 9x20mm screw and then with a 9x30mm screw.The broken pieces were retrieved from the patient.There was no harm for patient, operator or third party.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
 
Additional Manufacturer Narrative
Complaint is confirmed.One unpackaged ar-4020c-09 interference screw 9 x 20 mm serial/batch number (b)(6) was received for investigation.Visual evaluation found that the screw lost geometry on the thread, and it was broken off at the distal end at the tip.The cause remains undetermined.However, the most likely causes are improper bone prep, misaligned insertion, prying and/or leveraging the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Ø9X 20MM BC IF SCRW, VENTED
Common Device Name
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key16263368
Report Number1220246-2023-06175
Device Sequence Number7388886
Product Code MAI
UDI-Device Identifier00888867276970
UDI-Public00888867276970
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K180662
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2026
Device Model NumberØ9X 20MM BC IF SCRW, VENTED
Device Catalogue NumberAR-4020C-09
Device Lot Number15004895
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/10/2023
Supplement Date Received by Manufacturer01/10/2023
Initial Report FDA Received Date01/30/2023
Supplement Report FDA Received Date05/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-