| Model Number |
0502503030 |
| Medical Device Problem Code |
Fogging (1253)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/11/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that there was foggy image.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: foggy.Confirmed failure: outer tube damaged (bent, dented).Broken rod lens.Damaged distal cover glass.Probable root cause: laser welding seal failure.Distal/proximal window solder failure.Damage to optical train.Damage to needle.Damage to distal or proximal windows.Moisture intrusion.End of life wear-out.Environmental disturbance: endoscope colder than dew point.Environmental disturbance: fluid entrapment between the coupler and eyepiece junction (eyepiece only).Use error.The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|
Event or Problem Description
|
|
It was reported that there was foggy image.
|
| |
|