• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONMED UTICA FRAZIER, WITH CNTL VENT + OBTR, 12 FR.; APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONMED UTICA FRAZIER, WITH CNTL VENT + OBTR, 12 FR.; APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED Back to Search Results
Model Number 0033120
Medical Device Problem Code Delivered as Unsterile Product (1421)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
During incoming inspection, the distributor rejected this device, 0033120, for an insufficient heatseal.There was no contact with the patient as this was found during incoming inspection.This will be reported as a malfunction with potential for injury upon reoccurrence.
 
Additional Manufacturer Narrative
The device has not been returned to date and no photographic evidence was provided.Therefore, the reported event cannot be verified.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.A two-year review of complaint history revealed there has been a total of (b)(4) complaints, regarding (b)(4) devices, for this device family and failure mode.During this same time frame (b)(4) devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be (b)(4).This issue will continue to be monitored through the complaint system to assure patient safety.
 
Additional Manufacturer Narrative
Update: reported event of ¿insufficient heatseal & (sterile) pouch or blister pack¿ was not confirmed during the evaluation process.Received five 0033120 in unopened original packaging.Lot number was verified.Performed a visual inspection, there were no obvious signs of a breach in the seal however, the device was encroaching into the seal.Performed a functional inspection, the devices were dye leak tested per tm-10-217 rev.F which indicated that the packaging did not have an insufficient heat seal.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.A two-year review of complaint history revealed there has been a total of 28 complaints, regarding 55 devices, for this device family and failure mode.(b)(4).We will continue to monitor for trends through the complaint system to assure patient safety.
 
Event or Problem Description
During incoming inspection, the distributor rejected this device, 0033120, for an insufficient heatseal.There was no contact with the patient as this was found during incoming inspection.This will be reported as a malfunction with potential for injury upon reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FRAZIER, WITH CNTL VENT + OBTR, 12 FR.
Common Device Name
APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED
Manufacturer (Section D)
CONMED UTICA
525 french rd
utica NY 13502
Manufacturer (Section G)
CONMED UTICA
525 french rd
utica NY 13502
Manufacturer Contact
samantha dewberry
525 french road
utica, NY 13502
3152230184
MDR Report Key16266048
Report Number1320894-2023-00021
Device Sequence Number11805587
Product Code GCX
UDI-Device Identifier30653405000147
UDI-Public(01)30653405000147(17)260613(10)202106141
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 03/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date06/13/2026
Device Model Number0033120
Device Catalogue Number0033120
Device Lot Number202106141
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/17/2023
Supplement Date Received by Manufacturer02/20/2023
Initial Report FDA Received Date01/30/2023
Supplement Report FDA Received Date03/08/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
Patient EthnicityNon Hispanic
-
-