An olympus representative reported to olympus, on behalf of the customer, that the single use repositionable clip did not deploy after it had grabbed the tissue during an unknown procedure.The clip had to be forcibly removed, which caused a small tear in the mucosa.Attempts to obtain additional information was performed but no further information was available.The non-deployment of clips occurred 3 times; at least once resulted to a small mucosal tear while the other 2 instances did not.These incidents are captured under the following patient identifiers: (b)(6).
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device was not returned; therefore, the reported phenomenon or condition of the device could not be confirmed.The device history record was unable to be reviewed for this device since the serial number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.The specific root cause of the reported problem could not be determined at this time because the device was not returned.The following information is stated in the instructions for use (ifu): ¿·operation of this instrument is based on the assumption that open surgery is possible as an emergency measure if the clip cannot be detached from the instrument or if any other unexpected circumstance takes place.In this case, refer to chapter 14, ¿emergency treatment¿.·do not withdraw the instrument forcibly or change the angulation of the endoscope when the clip is grasping the tissue and is not released.Doing so may tear tissue inside the body cavity, resulting in patient injury, such as perforation, bleeding, or mucous membrane damage.·do not angulate the bending section of the endoscope or withdraw the instrument from the endoscope while the clip is grasping the tissue before being released.This could cause patient injury, such as perforation, bleeding, or mucous membrane damage.·do not try to forcibly withdraw the instrument from the endoscope if the clip cannot be detached from the instrument.Forcibly withdrawing the instrument could cause patient injury such as perforation, bleeding, or mucous membrane damage.·should the slightest irregularity and/or breakage of the parts be suspected, withdraw the instrument from the endoscope immediately according to the steps of ¿withdrawing the instrument from the endoscope¿ on page 10.Otherwise, perforation, bleeding, or mucous membrane damage may result.¿ olympus will continue to monitor field performance for this device.
|