| Model Number |
N/A |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
01/11/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient underwent a revision surgery of the left knee due to aseptic loosening and dislocation of mobile bearing.Due diligence is in progress for this complaint; to date no additional information or product has been received.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Concomitant medical products: oxf uni tib tray sz b lm pma item# 154720 lot# 960970.Oxf twin-peg cmntd fem sm pma item# 161468 lot# 591120.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 3002806535-2023-00025.3002806535-2023-00034.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: a2, b4, b5, d6, g3, g6, h1, h2, h10 if any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).No product was returned or pictures provided; visual and dimensional evaluations could not be performed.A review of the device manufacturing records confirmed no abnormalities or deviations.Devices are used for treatment.Radiographs were provided and reviewed by a radiologist.Significant findings include apparent femoral implant loosening and bearing dislocation with tibiofemoral subluxation as noted.Implant size appears appropriate and there is no radiographic evidence of trauma or injury.With the available information, a definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|