| Model Number |
US14150515 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Material Rupture (1546); Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/05/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an angioplasty procedure of a calcified target lesion in the superficial femoral artery, the pta balloon allegedly ruptured at the rated burst pressure.It was further reported that the guidewire allegedly got stuck upon removal.Reportedly, the balloon and the guidewire were removed together as one unit.The procedure was completed using another device.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
As the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiry date: 08/2024).
|
| |
|
Event or Problem Description
|
|
It was reported that during an angioplasty procedure of a calcified target lesion in the superficial femoral artery, the pta balloon allegedly ruptured at the rated burst pressure.It was further reported that the guidewire allegedly got stuck upon removal.It was further reported that sheath allegedly had difficulty in retracting the balloon through the introducer sheath.Reportedly, the balloon and the guidewire were removed together as one unit.The procedure was completed using another device.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.A definitive root cause for the reported balloon rupture, guidewire removal difficulty and sheath removal difficulty could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: b5, d4 (expiry date: 08/2024), g3, h6 (method).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|