• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND ONYX FRONTIER; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND ONYX FRONTIER; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number ONYXNG22526UX
Medical Device Problem Codes Device Dislodged or Dislocated (2923); Positioning Problem (3009)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/07/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product analysis: the device was received for analysis.A kink was evident to the distal shaft.The stent was not present on the balloon and did not return for analysis.The balloon folds remained intact.Crimp impressions were visible on the exposed balloon surface.Deformation was evident to the distal tip.The inner lumen patency was verified with a 0.015 inch mandrel.No other damage evident to the remainder of the device.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
An attempt was made to use one onyx frontier coronary drug eluting stent to treat a moderately tortuous, moderately calcified lesion, with 90% stenosis, in the ostium and distal left main (lm) coronary artery and circumflex (cx) artery.The device was inspected with no issues noted.Negative prep was performed with no issues noted.The lesion was pre-dilated.The device did pass through a previously deployed stent.Resistance was not encountered when advancing the device.Excessive force was not used.It was reported that the stent failed to cross the lesion.It was also reported that stent dislodgement occurred in vitro.The lesion was treated with rotational ablation using a non-medtronic (mdt) 1.75mm burr, as well as a 2.0 x 20mm non-mdt balloon and a 2.5 x 15mm non-mdt nc balloon.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ONYX FRONTIER
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key16272028
Report Number9612164-2023-00454
Device Sequence Number6033910
Product Code NIQ
UDI-Device Identifier00763000511234
UDI-Public00763000511234
Combination Product (Y/N)N
Initial Reporter StatePR
Initial Reporter CountryUS
PMA/510(K) Number
P160043
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/22/2024
Device Model NumberONYXNG22526UX
Device Catalogue NumberONYXNG22526UX
Device Lot Number0011034349
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/22/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/13/2023
Initial Report FDA Received Date01/31/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age67 YR
Patient SexMale
-
-