Model Number CYTO-201
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 01/13/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned.¿ these words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, a patient with the history of mild difficulties of dysphagia, during the cytosponge procedure, the nurse tried to retrieve the sponge by pulling on the string.The sponge had detached from the string which was located 30 cm from the mouth not far from the gastroesophageal junction.Endoscopy was performed in a hospital within 4 hours of procedure to retrieve the devices.The patient was prepped for the procedure.
 
Event or Problem Description
According to the reporter, during the procedure of a patient with the history of mild difficulties of dysphagia.The nurse tried to retrieve the sponge by pulling on the string.The sponge had detached from the string, which was located 30 cm from the mouth not far from the gastroesophageal junction within 5-10cm of the top of the stomach in the esophagus.No reported symptoms experienced by the patient after the detachment.The patient had a throat spray for the endoscopy to removed the device.Endoscopy was performed in a hospital within 4 hours of procedure to retrieve the devices.The patient was prepped for the procedure.
 
Additional Manufacturer Narrative
Additional information: g3, h6, h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the sponge detached from the suture.It was reported that the suture broke after the patient swallowed the sponge.The reported issue was confirmed.The most likely cause was traced to a component failure.Internal process improvements have been initiated to mitigate this issue.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank, because the information is currently, unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CYTOSPONGE
Common Device Name
ESOPHAGOSCOPE (FLEXIBLE OR RIGID)
Manufacturer (Section D)
MMJ SA DE CV (USD)
ave henequen no 1181 desarroll
ciudad juarez 32575
MX  32575
Manufacturer (Section G)
MMJ SA DE CV (USD)
ave henequen no 1181 desarroll
ciudad juarez 32575
MX   32575
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key16272456
Report Number9681384-2023-00007
Device Sequence Number20331370
Product Code EOX
UDI-Device Identifier10884521767782
UDI-Public10884521767782
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K181020
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/22/2023
Device Model NumberCYTO-201
Device Catalogue NumberCYTO-201
Device Lot Number22H0821JZX
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/16/2023
Supplement Date Received by Manufacturer02/13/2023
06/27/2023
Initial Report FDA Received Date01/31/2023
Supplement Report FDA Received Date02/16/2023
07/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age79 YR
Patient SexMale
-
-