| Model Number |
N/A |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/12/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4) customer has indicated that the product will be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that an instrument fell and fractured during verification.There is no additional information available.
|
| |
|
Additional Manufacturer Narrative
|
|
No product was returned, or pictures provided; therefore, visual and dimensional evaluations could not be performed.Reported event did not occur in an operating room or as part of a medical procedure; medical records are not available for review.Device history record (dhr) was reviewed for deviations and/ or anomalies with no deviations / anomalies identified.The root cause of the reported issue is attributed to user mishandling the device.Per package insert 87620399923 states not to subject instruments to high loads and/or impact as breakage can occur.If any further information is received which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Search Alerts/Recalls
|