Model Number VOCSN-VC PRO, ENGLISH
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/30/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the device needed maintenance.Upon evaluation of the device, ventec observed that the ventilator was unable to maintain peep (positive end expiratory pressure).There was no patient involvement associated with the reported event.
 
Additional Manufacturer Narrative
The device was evaluated by ventec where the reported issue of the ventilator being unable to maintain peep (positive end expiratory pressure) was confirmed.Ventec replaced the exhalation control module (ecm) to resolve the reported issue.Proper device operation was then observed through functional and performance testing.The investigation determined that the root cause of the reported issue was the ecm.
 
Additional Manufacturer Narrative
Corrected information for initial medwatch report: section d4, model #, of the initial medwatch report stated: prt-00900-001.Section d4, model #, of the initial medwatch report should have stated: vocsn-vc pro, english.Section d4, udi #, of the initial medwatch report stated: (b)(4).Section d4, udi #, of the initial medwatch report should have stated: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VOSCN
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
VENTEC LIFE SYSTEMS, INC
22002 26th ave se
bothell WA 98021
Manufacturer (Section G)
VENTEC LIFE SYSTEMS, INC
22002 26th ave se
bothell WA 98021
Manufacturer Contact
mike kubany
22002 26th ave se
bothell, WA 98021
4256861765
MDR Report Key16273141
Report Number3013095415-2023-00045
Device Sequence Number20705935
Product Code CBK
UDI-Device Identifier00855573007433
UDI-Public(01)00855573007433(11)190104
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K162877
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVOCSN-VC PRO, ENGLISH
Device Catalogue NumberPRT-00900-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/14/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/14/2022
Supplement Date Received by Manufacturer08/23/2025
Initial Report FDA Received Date01/31/2023
Supplement Report FDA Received Date08/23/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/04/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-