• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS VASCULAR GUIDEWIRE; VASCULAR GUIDEWIRE, NITINOL W/ HYDROPHILIC COATING, STRAIGHT, RADIOPAQUE TIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS VASCULAR GUIDEWIRE; VASCULAR GUIDEWIRE, NITINOL W/ HYDROPHILIC COATING, STRAIGHT, RADIOPAQUE TIP Back to Search Results
Catalog Number 45747
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 01/20/2023
Type of Reportable Event Malfunction
Event or Problem Description
A territory manager reported an end user experienced a procedure in which they had an issue using a guidewires, nitinol.018x145cm, tungsten coil, hydrophilic kit.During the procedure, it was reported the tip of the guidewire detached inside the patient.The tip was successfully retrieved; however, the procedure was not continued.The patient did not experience any adverse effects or harm due to this incident.
 
Additional Manufacturer Narrative
The reported device has yet to be returned to the manufacturer for a device evaluation.An investigation into the root cause for product problem is currently in progress.The results of the investigation will be sent via a follow up medwatch.Reference (b)(4).
 
Additional Manufacturer Narrative
As the reported device was not returned, angiodynamics is unable to perform a device evaluation.The customer's reported complaint description of guidewire tip detached inside the patient was not confirmed; the guidewire sample was not returned.Without receiving the complaint sample for evaluation, a definitive root cause cannot be determined.Potential root cause for this type of guidewire failure mode is the end user pulling the guidewire back against the needle if resistance to advancement is observed.Dhr review of the indicated packaging lot showed no device non-conformance at the time of device distribution.The guidewire in question is a finished packaged good that is purchased from heraeus medical.Heraeus medical was made aware of this reported event via fyi scar00004741.A review of the device history records was performed for the indicated lots for any deviations related to the reported failure mode of the complaint.The review confirms that the lots met all material, assembly and performance specifications; i.E.No ncr associated with reported failure mode.Labeling review: dfu provided with this guidewire device contains the following statements: directions: insert the guidewire into the needle and advance the guidewire to the targeted location.The guidewire tip can be visualized under fluoroscopy during insertion.Once the guidewire is inserted, remove the needle.Do not advance the guidewire against resistance until the cause of the resistance has been determined.Cautions: do not advance the vascular guidewire against resistance until the cause of the resistance has been determined.Excessive force against resistance may result in damage to the guidewire and catheter or vessel perforation.Navilyst medical, inc.Does not recommend a particular technique for the use of this device.The steps contained in the preceding directions discuss the seldinger technique for percutaneous entry and are of information purposes only.Each physician should evaluate their appropriateness according to individual patient conditions and his or her medical training and experience.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VASCULAR GUIDEWIRE
Common Device Name
VASCULAR GUIDEWIRE, NITINOL W/ HYDROPHILIC COATING, STRAIGHT, RADIOPAQUE TIP
Manufacturer (Section D)
ANGIODYNAMICS
10 glens falls technical park
glens falls NY 12801
Manufacturer (Section G)
ANGIODYNAMICS
10 glens falls technical park
glens falls NY 12801
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
5086587990
MDR Report Key16273419
Report Number1317056-2023-00013
Device Sequence Number7455392
Product Code DQX
UDI-Device IdentifierH965457471
UDI-PublicH965457471
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K070150
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number45747
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/20/2023
Supplement Date Received by Manufacturer05/04/2023
Initial Report FDA Received Date01/31/2023
Supplement Report FDA Received Date05/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-