| Model Number |
OJR410 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/26/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Fisher & paykel healthcare (f&p) are currently in the process of retrieving further information regarding this complaint.We will provide a follow up report upon the completion of our investigation.
|
| |
|
Event or Problem Description
|
|
A distributor in japan on behalf of a healthcare facility reported via a fisher and paykel (f&p) healthcare field representative that the tubing of an ojr410 optiflow junior 2 nasal cannula was detached from the cannula tube joint during patient use.There was no reported patient consequence.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The optiflow junior cannula is designed specifically for the delicate anatomical features and flow requirements of neonatal and paediatric patients.It features adhesive pads (wigglepads) to maintain cannula stability on the patient's cheeks, soft-touch nasal prongs and breathable kink-proof and crush-resistant flexible tubing.Method: the complaint device ojr410 optiflow junior 2 nasal cannula was received at fisher & paykel healthcare (f&p) in new zealand for evaluation where it was visually inspected.Results: visual inspection of the complaint cannula revealed that the right side of the tubing was detached from the cannula.It was also noted that the glue residue was visible on the surface of the detached tube end and inside the cannula tube joint.It was also noted that glue residue was visible inside both tubes near the cannula.Conclusion: we are unable to determine the cause of the damage.However, the observed damage was most likely caused by the tubing being pulled.All optiflow junior cannulas are 100% leak and occlusion tested after final assembly and any cannula that fails is discarded.Samples are also taken and pull tested to check the tube tensile strength and the glue joint strength at the cannula/tube joint, as well as the swivel grip joint.The subject cannula would have met the required specification at time of production.The user instructions illustrate in pictorial format the correct set-up and proper use of the optiflow junior 2 nasal cannula.They also state the following: appropriate patient monitoring (e.G.Oxygen saturation) must be used at all times.Failure to monitor the patient (e.G.In the event of an interruption to gas flow) may result in serious harm or death).Do not wrap, insulate, stretch or crush the tubing as this may impair the performance of this product or compromise safety (including potentially causing patient harm).Ensure all connections are secure during use.Check the cannula is undamaged and that the flow path is maintained.Under excessive load, the cannula may disconnect to prevent forces being transferred to the patient.
|
| |
|
Event or Problem Description
|
|
A distributor in japan on behalf of a healthcare facility reported via a fisher and paykel (f&p) healthcare field representative that the tubing of an ojr410 optiflow junior 2 nasal cannula was detached from the cannula tube joint during patient use.There was no reported patient consequence.
|
| |
|