• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVANOS MEDICAL INC. BALLARD CLOSED SUCTION SYSTEM NEO/PED 7FR; VAP CLOSED SUCTION CATHETERS & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. BALLARD CLOSED SUCTION SYSTEM NEO/PED 7FR; VAP CLOSED SUCTION CATHETERS & ACCESSORIES Back to Search Results
Model Number 197
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Decreased Peak Expiratory Flow rate (2435)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The actual complaint product was not returned for evaluation.A review of the device history record is in-progress.All information reasonably known as of 31-jan-2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.Device not returned.
 
Event or Problem Description
Avanos medical received a single report that referenced two different incidences, which were associated with separate units, involving two different events.This is the first of two reports.Refer to 8030647-2023-00012 for the second event.It was reported "a break in the plastic sheath covering [the] inline ballard suction catheter result[ed] in loss of ventilator pressure to the infant." there was no reported injury.Additional information received 11-jan-2023 stated the closed suction catheter (csc) was not altered in anyway prior to the reported incident occurring.
 
Additional Manufacturer Narrative
The device history record for the reported lot number, 30219584, was reviewed and documented that the lot was manufactured according to the approved manufacturing procedures, specifications, and then released by quality assurance.All information reasonably known as of 05-apr-2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BALLARD CLOSED SUCTION SYSTEM NEO/PED 7FR
Common Device Name
VAP CLOSED SUCTION CATHETERS & ACCESSORIES
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT S DE RL DE CV
carretera internacional
salida norte no. 1053
magdalena, cp
Manufacturer Contact
lisa clark
5405 windward parkway
alpharetta, GA 30004
4704485444
MDR Report Key16274666
Report Number8030647-2023-00011
Device Sequence Number7492321
Product Code BSY
UDI-Device Identifier00609038938257
UDI-Public00609038938257
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number197
Device Catalogue NumberN/A
Device Lot Number30219584
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/04/2023
Supplement Date Received by Manufacturer03/09/2023
Initial Report FDA Received Date01/31/2023
Supplement Report FDA Received Date04/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/14/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-