Model Number TABLETOP |
Device Problems
Defective Component (2292); Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
A customer reported that in an ophthalmic console the foot pedal was coming apart and both illuminator lights are out.Procedure details and patient impact were not reported.
|
|
Manufacturer Narrative
|
The company representative was able to replicate the reported event.The nonconforming footswitch and lamp were both replaced to address the issue.The system was tested and found to meet product specifications.A non-conformance based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this serial number was performed.A potentially relevant complaint was found and reviewed as part of this investigation.The root cause of the reported event is attributed to nonconforming lamp and a nonconforming footswitch.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
|
|