Catalog Number 51-104130
Medical Device Problem Code Manufacturing, Packaging or Shipping Problem (2975)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/06/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Report source: japan.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2023-00178 and 0001825034-2023-00180.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported while inspecting the items in stock, the sterile packing was found to be damaged.There was not patient involvement.Attempts have been made and no further information is available.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.Visual evaluation of the returned product identified damage to the sterile packaging (blister).Sterility has not been compromised.Reported event has been confirmed by review of the returned product.Dhr was reviewed and no discrepancies related to the reported event were found.Investigation results concluded that the reported event was due to transit damage and a packaging design issue.The condition of the device when it left zimmer biomet is conforming to specification.The device evaluation found the product to be conforming and sterility not compromised.Further, the event did not contribute to a serious injury; therefore, this would not be considered a reportable event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No additional event information to report at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC 133 T1 PPS HO 13X146MM
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key16276477
Report Number0001825034-2023-00179
Device Sequence Number6390856
Product Code LZO
UDI-Device Identifier00880304489738
UDI-Public(01)00880304489738(17)280430(10)6278597
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K200196
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number51-104130
Device Lot Number6278597
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/06/2023
Supplement Date Received by Manufacturer04/04/2023
Initial Report FDA Received Date02/01/2023
Supplement Report FDA Received Date06/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/30/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-