• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AOMORI OLYMPUS CO., LTD. SINGLE USE INJECTOR; INJECTOR AND SHEATHSET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AOMORI OLYMPUS CO., LTD. SINGLE USE INJECTOR; INJECTOR AND SHEATHSET Back to Search Results
Model Number NM-401L-0423
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/29/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported that a therapeutic gastric ulcer bleeding case was being done.Prior to the procedure, two injectors were opened and inspected by the nurse.No issues were found when the injectors were tested.During the procedure, when the injector, with the needle out, was inside the scope, the nurse injected the adrenaline solution; however, the solution was unable to flow.The solution was stuck and could not flow through the lumen despite multiple attempts.The nurse tried to examine and test the injector after taking it out of the scope, but the same issue occurred.The procedure was completed using another injector.There was no reported patient harm or impact due to this event.
 
Additional Manufacturer Narrative
Reports are being submitted on both single use injectors used during the procedure.Please refer to the following reports: patient identifier of (b)(6) is related to model number: nm-401l-0423, serial number: (b)(4).Patient identifier of (b)(6) is related to model number: nm-401l-0423, serial number: (b)(4).During device evaluation at olympus, it was found the complaint was confirmed and the tube was kinked.In addition, evaluation found the needle was unable to extend and be retracted (able to inject).This event is under investigation.A supplemental report will be submitted upon receiving additional information.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it was likely that the phenomenon occurred due to the compressive bucking on the needle tube in the handle.The compressive buckling on the needle tube was likely caused when the needle was extended because of the great friction between the outer tube and the needle.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE INJECTOR
Common Device Name
INJECTOR AND SHEATHSET
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16276509
Report Number9614641-2023-00145
Device Sequence Number17743823
Product Code FBK
UDI-Device Identifier04953170422621
UDI-Public04953170422621
Combination Product (Y/N)N
Initial Reporter CountryMY
PMA/510(K) Number
K902736
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberNM-401L-0423
Device Lot Number23K
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/10/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/08/2023
Supplement Date Received by Manufacturer02/06/2023
Initial Report FDA Received Date02/01/2023
Supplement Report FDA Received Date03/02/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/06/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-