Model Number A209
Device Problems No Audible Alarm (1019); Interrogation Problem (4017)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/26/2023
Event Type  Injury  
Event Description
It was reported that, during an office follow-up, this device could not be interrogated.The application of a magnet did not yield tones.Technical services suspects the battery has reached end-of-life.The device has been implanted greater than five years.The device remains implanted.There were no adverse patient effects.
 
Event Description
This supplemental report is being filed to provide additional information that the product has been explanted and replaced.There were no additional adverse patient effects.It was reported that, during an office follow-up, this device could not be interrogated.The application of a magnet did not yield tones.Technical services suspects the battery has reached end-of-life.The device has been implanted greater than five years.The device remains implanted.There were no adverse patient effects.
 
Event Description
This supplemental report is being filed to provide additional information regarding product analysis.It was reported that, during an office follow-up, this device could not be interrogated.The application of a magnet did not yield tones.Technical services suspects the battery has reached end-of-life.The device has been implanted greater than five years.The device remains implanted.There were no adverse patient effects.
 
Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, this s-icd was thoroughly inspected and analyzed.A high current drain was detected, but the battery still supported full device function.Detailed analysis confirmed the identified high current drain was a result of a compromised capacitor.This resulted in the observed premature battery depletion.It was determined that the capacitor was compromised due to the presence of excess hydrogen in the device case.Boston scientific issued an advisory communication in august 2019 regarding a subset of emblem devices that has an elevated potential of exhibiting this behavior; the population of devices that may be impacted was expanded in december 2020.This particular device is included in the emblem s-icd accelerated depletion advisory population.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMBLEM S-ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key16277929
MDR Text Key308558376
Report Number2124215-2023-04098
Device Sequence Number1
Product Code LWS
UDI-Device Identifier00802526544101
UDI-Public00802526544101
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P110042/S043
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup,Followup
Report Date 08/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/01/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/30/2017
Device Model NumberA209
Device Catalogue NumberA209
Device Lot Number109271
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/22/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/30/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-0935-2021
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age57 YR
Patient SexFemale
-
-