| Model Number |
71654501 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/04/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, during internal fixation surgery, the nail did not engage well in a one (1) std drill guide and the graft of the bow did not allow easy insertion.In fact, to remove it, it split.It is unknown if pieces fell into the patient.It is unknown how surgery was completed an if there was a delay.
|
| |
|
Event or Problem Description
|
|
It was reported that, during internal fixation surgery, one (1) trigen meta anterior drop did not fit well into a trigen standard drill guide, to which the nailed was anchored and the graft of the bow did not allow easy insertion.In fact, to remove it, the trigen meta anterior drop split.It is unknown if pieces fell into the patient.It is unknown how surgery was completed an if there was a delay.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.However, the photographs were reviewed, and revealed that the device is fractured.Device batch number was not provided, thus, an evaluation of the manufacturing records could not be performed.A review of complaint history of the previous 12 months revealed similar events for the listed device, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Internal complaint reference number: case-(b)(4).Section d4 was corrected.
|
| |
|