| Model Number |
IS3-WW |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Loss of Vision (2139)
|
| Date of Event |
10/26/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that following a left eye (os) trabecular microbypass system procedure, the patient presented at one (1) month postop with "loss of visual acuity (va) of two (2) or more lines from the baseline".Additional information is being requested.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not available; therefore, product testing on the actual device could not be performed.The device history record review of the manufacturing lot was performed and there were no non-conformities found to be related to the reported event.A review of the device labeling was completed.Decreased/impaired vision is identified in the labeling as a known inherent risk of trabecular micro-bypass stent procedure.The ifu adequately provides instructions for stent implantation, precautions, and warnings for the proper use and handling of the device.An adverse event appears to have occurred, but does not appear to have been a problem with the device or the way it was used.Decreased/impaired vision is an established risk associated with use of the device, which is clearly specified in the product''s labeling.Based on the information received, the root cause of the reported event could not be conclusively identified.Mfr# reference: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Mfr# reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
Through follow-up, the patient''s laboratory results associated with visual acuity were received.
|
| |
|