| Model Number |
VCP490G |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/18/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.When was the needle detached from the thread (in the package, during removal from package, during handling prior to use on patient or during use on the patient)? please specify.Photo analysis summary: this is an analysis for a photo submitted to ethicon for evaluation.During the visual analysis, the following was observed: the photo shows an open vcp490g product box with three open aluminum packages.The image is not clear to determine the failure mode or the reported condition.Based on the photo review, the event describe is not confirmed, and no conclusion or root cause could be determined.Hands on device analysis may provide the additional evidence necessary to confirm the root cause of the reported event.Because the instrument was not returned our evaluation is limited.As part of ethicon quality process all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.Attempts are being made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Component code: g07002.A manufacturing record evaluation was performed for the finished device batch, and no non-conformances were identified.Trade name - irgacare®.Active ingredient(s) ¿ triclosan.Dosage form ¿ suture/solid/parenteral.Strength ¿ = 275 g/m.Related events captured via 2210968-2023-00707, 2210968-2023-00708.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent an unknown procedure on (b)(6) 2023 and suture was used.The needle detaches from the thread just like control release sutures.There were no adverse patient consequences reported.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 2/23/2023.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 no device problem found.Additional information was requested, the following was obtained: according to the customer (head nurse), the needle detached from the thread during handling prior to use on patient and during use on the patient h3 investigational summary: the product was returned incomplete to ethicon for evaluation.An analysis of the product could not be performed since a physical sample was not received for evaluation, only one empty opened foil.A manufacturing record evaluation was performed for the finished device, and no non-conformances related to the reported complaint condition were identified.As part of ethicon's quality process, each batch is randomly visually inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.If the product is received at a later date, the investigation will be updated as applicable.
|
| |
|
Search Alerts/Recalls
|
|
|