• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT CURRENT PLUS VR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT CURRENT PLUS VR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD1211-36
Medical Device Problem Code Premature Discharge of Battery (1057)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/30/2022
Type of Reportable Event Serious Injury
Event or Problem Description
During an in clinic follow up, upon interrogation the longevity was less than expected from the previous follow up 7 months ago.Premature battery depletion was suspected.The device was explanted to resolve the event.The patient was stable.
 
Additional Manufacturer Narrative
Further information was requested but not received.
 
Additional Manufacturer Narrative
Upon receipt, the device was at elective replacement indicator (eri).A longevity analysis was performed using available programmed settings and usage data and found the battery depletion to be normal.The device functioned normally on the bench and no anomalies were found.Premature depletion was not confirmed by analysis.During data review of the device image, analysis found the battery life estimated on the programmer was higher than the actual battery life remaining.This longevity overestimation was determined to be due to an estimation inaccuracy in the merlin programmer software.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CURRENT PLUS VR
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16279500
Report Number2017865-2023-04735
Device Sequence Number17906895
Product Code NVZ
UDI-Device Identifier05414734503686
UDI-Public05414734503686
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
P910023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2015
Device Model NumberCD1211-36
Device Catalogue NumberCD1211-36
Device Lot Number4418271
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/16/2023
Supplement Date Received by Manufacturer02/15/2023
03/29/2023
Initial Report FDA Received Date02/01/2023
Supplement Report FDA Received Date02/17/2023
04/04/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/12/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-