| Model Number |
CD1211-36 |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/30/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
During an in clinic follow up, upon interrogation the longevity was less than expected from the previous follow up 7 months ago.Premature battery depletion was suspected.The device was explanted to resolve the event.The patient was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Further information was requested but not received.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt, the device was at elective replacement indicator (eri).A longevity analysis was performed using available programmed settings and usage data and found the battery depletion to be normal.The device functioned normally on the bench and no anomalies were found.Premature depletion was not confirmed by analysis.During data review of the device image, analysis found the battery life estimated on the programmer was higher than the actual battery life remaining.This longevity overestimation was determined to be due to an estimation inaccuracy in the merlin programmer software.
|
| |
|