• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER TRAUMA KIEL NAIL HOLDING SCREW GAMMA3® 8X35MM; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER TRAUMA KIEL NAIL HOLDING SCREW GAMMA3® 8X35MM; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 1320-0090
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/08/2023
Type of Reportable Event Malfunction
Event or Problem Description
As reported: "the end user reported that gamma3 nail can no longer be disconnected from the nail holding screw and the aiming device." completely had to be removed, new implant had to be placed freehand and the procedure took longer, leading to an unknown surgical delay.
 
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event or Problem Description
As reported: "the end user reported that gamma3 nail can no longer be disconnected from the nail holding screw and the aiming device." completely had to be removed, new implant had to be placed freehand and the procedure took longer, leading to an unknown surgical delay.
 
Additional Manufacturer Narrative
Please note the correction to h6 health impact codes.The reported event could be confirmed, since the product was returned for evaluation and matches the alleged failure mode.The device inspection revealed the following: the received device came back jammed/assembled which was unblocked later.The device shows signs of usage, but no visible major damage was observed which can affect the functionality of the device.However dried blood was observed inside and outside on the surface of the screw.Post-decontamination and after cleaning dried blood, a pre-surgical test was conducted for the returned devices.Upon inspection, the devices were functionally ok.Assembling and disassembling of the nail-holding screw can be easily done through the target device with the nail without any resistance.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Based on the investigation, the root cause of the issue is most likely user related.The devices shows the appearance of dried blood residual on their surfaces, which potentially caused overtightening of the nail in between the interface of the nail, target device and the screw.If any further information is provided, the complaint report will be updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAIL HOLDING SCREW GAMMA3® 8X35MM
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key16280458
Report Number0009610622-2023-00020
Device Sequence Number12392729
Product Code LXH
UDI-Device Identifier04546540377661
UDI-Public04546540377661
Combination Product (Y/N)N
Initial Reporter CountryBE
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 03/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1320-0090
Device Catalogue Number13200090
Device Lot NumberKSS13N227
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/31/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/08/2023
Supplement Date Received by Manufacturer02/28/2023
Initial Report FDA Received Date02/01/2023
Supplement Report FDA Received Date03/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/14/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-