| Model Number |
11522558 |
| Medical Device Problem Codes |
Leak/Splash (1354); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/30/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the bd alaris¿ pump module smartsite¿ infusion set tubing broke off below the cassette after infusing remicade and leaked.This complaint was created to capture the 2nd of 4 related incidents.The following information was provided by the initial reporter: "bd alaris ball valve pump infusion set was used for medication infusion (remicade).At the end of the infusion when attempting to remove the tubing from the pump for disposal, the tubing broke off below cassette.There have been four events in the past week where the tubing becomes disconnected from the cassette.In two cases the medication leaked through and had to be reordered and a second dose prepared by pharmacy.".
|
| |
|
Additional Manufacturer Narrative
|
|
Medical device expiration date: unknown.Device evaluated by mfr: a device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: investigation summary: a complaint of four events of separation were received from the customer.Pictures of the lot and the sample were returned for investigation.In the photo, the tubing is seen to be separated at the lower part of the pumping segment, after the safety clamp.The customer complaint of separation was confirmed.A device history review was completed for the known lot.A device history record review for model 11522558 lot number 22095036 was performed.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.After investigation, it was determined by the manufacturing plant that the most potential root cause that generates the separation of the tubing and lower fitment is the omission of solvent application by operator.In addition, although the process has an inspection step for 100% of the product, there were found opportunities in the method of inspection.A trend for this separation issue has been identified for this product line.A capa (corrective action preventative action) has been initiated, and a team has been assembled in order to investigate the issue.
|
| |
|
Event or Problem Description
|
|
It was reported that the bd alaris¿ pump module smartsite¿ infusion set tubing broke off below the cassette after infusing remicade and leaked.This complaint was created to capture the 2nd of 4 related incidents.The following information was provided by the initial reporter: "bd alaris ball valve pump infusion set was used for medication infusion (remicade).At the end of the infusion when attempting to remove the tubing from the pump for disposal, the tubing broke off below cassette.There have been four events in the past week where the tubing becomes disconnected from the cassette.In two cases the medication leaked through and had to be reordered and a second dose prepared by pharmacy.".
|
| |
|